Operating Instructions199255 COMPAT DualFlo®Enteral Delivery Pump199255 Manual 7/16/08 10:05 AM Page 2
9Detailed Operating InstructionsCheck pump settings1. Press ON- alarm sounds briefly and displays .8.8.8.8. The pump has memoryand displays previous s
10Detailed Operating Instructions (cont.)Formula Strength (dilution)5. To dilute formula, press SELECTuntil strength is displayed. Press orto set
11Detailed Operating Instructions (cont.)Begin Feeding1. Connect delivery set adapter to thepatients enteral feeding tube. CAUTION: Verify enteral tub
12Quick Reference(How to set Flow Rate, Dose Limit, or CLEAR Formula Volume Delivered)To clear Volume Delivered and/orAccumulated VolumeIf Formula Str
13Quick Reference(How to set Hydration with Automatic Flush)Select Time Interval (in hours) overwhich water volume will be delivered4. Press SELECT un
14Quick Reference(How to set Formula Strength, Programmable Flush, or CLEAR Water Volume Delivered)Programmable flush after dose isdeliveredA set volu
15Alarms/Indicatorsno SEt (No set)An audible alarm sounds, and no SEt isdisplayed. Indicates one of the followingconditions:• A COMPAT DualFlo set is
16Alarms/Indicators (cont.)HoLd Err (Hold error)Pump is left on hold and START/HOLDis not pressed within 21/2 minutes. PressSTART/HOLD to silence the
17TroubleshootingCOMPAT®Enteral Feeding Pumps are reliable electromechanical devices. As with anymechanical device, problems may occur. The following
18Specifications• 199407 COMPAT DualFlo®PiercingSpike and 1000 mL Vinyl Water Bagwith preattached pump set• 199447 COMPAT DualFlo®1000 mL Formula Viny
1Table of ContentsSection PageIntroduction 2Controls 4Preparation of COMPAT DualFlo®Delivery Set 8Operating Instructions 9Quick Reference:Flow Rate, D
19Specifications (cont.)Type of Device . . . . . . . . . . . . . . . . .Volumetric enteral feeding pumpMechanism . . . . . . . . . . . . . . . . . .
Symbols on Pump and Accessorries20The following symbols may be found on the pump:Refer to Accompanying DocumentsAlternating CurrentType BF Protection
21Self Test5. The pump rotor should begin toturn. To test the visual alarms, presseither SELECT, CLEAR, , ,PRIME, or START/HOLD. Whenany of t
22Limited Warrantyof Nestlé HealthCare Nutrition.Nestlé HealthCare Nutrition reserves theright to repair or replace (at its option)any pump which fail
CMPT-10267-0308SN1032199255 Manual 7/16/08 10:05 AM Page 1
IntroductionDelivery Set Security Doors Reduces the likelihood of accidentaldislodgment of sets and enablesimmediate detection of a free flowcondition
3Lifetime memory retention Infusion Rate, Dose Limit, FormulaStrength, Water Volume Set, TimeInterval, Formula and Water VolumeDelivered, and Accumula
4ControlsFunction Displays9. RUNNING, STOPPED, indicates status of the pump when power is on.10. AC, indicates the unit is plugged intoan electrical o
5Controls (cont.)Front of the pump13. Flow Sensors and Drip Chamber Guide14. Tubing Guide15. Drip Chamber Brackets16. Rotary Peristaltic Pump Mechanis
6Controls (cont.)242526272829303132199255 Manual 7/16/08 10:05 AM Page 14
7Controls (cont.)Water Side Displays29. VOLUME SET, displays the volumeof water to be delivered over theselected Time Interval or at the endof the dos
8Preparation of COMPAT DualFlo®Delivery SetDIAGRAM 1Open pump doors by pulling door latch,insert pump set drip chamber into dripchamber bracket.DIAGRA